Weight 18 Kg Infusia Vp7 Volumetric Pump, Ambient Temperature Range 5 Degree Celsius To 40 Degree Celsius
Key Highlights
Available Variants
Technical Specifications
| Parameter | Value |
|---|---|
| Fuse | F1AL/250V, 2 PC Inside |
| Rate | 200ml/h |
| Brand | Byond |
| Weight | 1.8 Kg |
| Display | LCD |
| Voltage | 240V |
| KVO Rate | 5 ml/h (1ml/h step) |
| Warranty | 1 Year |
| Dimension | 140x157x220 mm (LWH) |
| Frequency | 50 Hz |
| Purge Rate | 1000ml/h (100 ml/h step) |
| Infusion Mode | Rate Mode |
| Occlusion Low | 300mmHg +/-100mmHg (40.7kPa +/-13.3kPa ) |
| Is It Portable | Portable |
| Occlusion High | 800mmHg +/- 200mmHg (106.7kPa +/- 26.7kPa) |
| Bubble Detector | Ultrasonic Wave Detector; Detection Sensitivity 25mu/l |
| Infusion Volume | 1ml |
| Battery Capacity | 1600mAh |
| Occlusion Medium | 500mmHg +/- 100mmHg (66.7kPa +/-13.3kPa) |
| IP Classification | IPX4 |
| Model Name/Number | BYS 820 |
| Power Consumption | 18VA |
| Relative Humidity | 90% |
| Usage/Application | A Bubble Detector And Pressure Sensor That Can Precisely Detect Air Bubbles |
| Flow Rate Accuracy | Within +/- 5% (After Correct Calibration) |
| Atmospheric Pressure | 106.0KPA |
| Mechanical Precision | Within +/- 2% |
| Minimum Order Quantity | 4 |
| Internal Battery Backup | 4 Hours |
| Equipment Classification | Class II |
| Ambient Temperature Range | +5 Degree Celsius To +40 Degree Celsius |